A Real-time Endoscopic Image Denoising System

TL;DR

Proposed a real-time endoscopic image denoising system, improving PSNR from 21.16 to 33.05.

eess.IV 🔴 Advanced 2025-06-18 36 views
Yu Xing Shishi Huang Meng Lv Guo Chen Huailiang Wang Lingzhi Sui
endoscope image denoising FPGA Poisson-Gaussian noise real-time processing

Key Findings

Methodology

The study proposes a hybrid denoising system combining traditional image processing algorithms with advanced learning techniques. Through physical modeling, it identifies three primary noise types: fixed-pattern noise, periodic banding noise, and mixed Poisson-Gaussian noise. The system is implemented on FPGA to ensure real-time performance.

Key Results

  • Experimental results show an average PSNR improvement from 21.16 to 33.05 on the test dataset, significantly reducing image noise while maintaining detail and color integrity.
  • After removing periodic banding noise and fixed-pattern noise, PSNR increased to 30.89 and 31.83, respectively.
  • Using the learning model, PSNR further improved to 33.05, close to the performance of CycleISP and SwinIR.

Significance

This study provides a practical denoising solution for medical endoscope systems, addressing the limited dynamic range and noise issues of small sensors, enhancing image quality and diagnostic reliability. The system can be directly integrated into existing image signal processing pipelines, with broad clinical application potential.

Technical Contribution

Technical contributions include: 1) a physics-based noise model covering multiple noise types; 2) a denoising system that does not require real paired data; 3) real-time processing capability on FPGA platforms.

Novelty

This is the first to propose a denoising system for medical endoscopes that combines physical modeling with learning methods, capable of handling complex noise types and achieving real-time performance.

Limitations

  • The system's performance in extreme low-light conditions needs further validation, as it may result in detail loss.
  • Limited adaptability to different sensor models, requiring additional calibration steps.

Future Work

Future work includes optimizing system performance under extreme conditions, extending to other types of medical imaging devices, and exploring more efficient hardware implementations.

AI Executive Summary

The miniaturization of endoscopes has greatly enhanced operational flexibility and diagnostic capability but also introduced image noise issues. Existing solutions fall short in addressing the noise from small sensors.

This paper proposes a hybrid denoising system combining traditional image processing algorithms with advanced learning techniques, effectively reducing noise in endoscopic images. Through physical modeling, it identifies three primary noise types and achieves real-time performance on FPGA platforms.

Experimental results show an average PSNR improvement from 21.16 to 33.05 on the test dataset, significantly reducing image noise while maintaining detail and color integrity. This study provides a practical denoising solution for medical endoscope systems, with broad clinical application potential.

Deep Analysis

Background

The miniaturization of endoscopes allows direct visualization of inaccessible tissues through narrow anatomical lumens, revolutionizing clinical diagnostics and biological research. However, the small sensor size limits photon capture, leading to significant image noise issues. Existing image denoising methods are limited in handling noise from small sensors, especially in high-contrast medical imaging scenarios.

Core Problem

Small sensors have a constrained dynamic range, making it difficult to capture details in both highlights and shadows. Additionally, simplified circuit design and analog signal transmission introduce extra noise sources, resulting in significant noise issues in processed endoscopic images. These issues negatively impact image quality and diagnostic accuracy.

Innovation

This paper innovatively proposes a denoising system combining physical modeling with learning methods. By analyzing a physics-based noise model, it identifies three primary noise types and develops a denoising system without requiring real paired data. The system can handle complex noise types and achieve real-time performance on FPGA platforms.

Methodology

  • �� Physics-based noise modeling: Identifies fixed-pattern noise, periodic banding noise, and mixed Poisson-Gaussian noise.

  • �� Denoising system design: Combines traditional image processing algorithms with learning techniques to process raw frames from sensors.

  • �� Real-time performance implementation: Achieves real-time processing capability on FPGA platforms.

Experiments

Experiments were conducted on two OV6946 sensors, capturing 200 sets of paired data as the test dataset. After removing periodic banding noise and fixed-pattern noise, PSNR increased to 30.89 and 31.83, respectively. Using the learning model, PSNR further improved to 33.05, close to the performance of CycleISP and SwinIR.

Results

Experimental results show an average PSNR improvement from 21.16 to 33.05 on the test dataset, significantly reducing image noise while maintaining detail and color integrity. After removing periodic banding noise and fixed-pattern noise, PSNR increased to 30.89 and 31.83.

Applications

The system can be directly integrated into existing image signal processing pipelines, enhancing endoscopic image quality and diagnostic reliability, with broad clinical application potential.

Limitations & Outlook

The system's performance in extreme low-light conditions needs further validation, as it may result in detail loss. Limited adaptability to different sensor models, requiring additional calibration steps.

Plain Language Accessible to non-experts

Imagine you're in a kitchen cooking. An endoscope is like a tiny kitchen camera used to observe the food in the pot. But because the camera is so small, there's not enough light, and the image has a lot of noise. This method is like a smart filter that removes the noise, allowing you to see the food's condition more clearly. The filter not only removes noise but also retains the food's color and details, just like using a high-quality filter in the kitchen to take pictures.

ELI14 Explained like you're 14

Imagine you're playing a game, and the screen is full of noise, affecting your gaming experience. This method is like a super cool game filter that removes the noise, allowing you to see the game screen more clearly. The filter not only removes noise but also retains the game's color and details, just like using a high-quality filter in the game to take pictures. Isn't that cool?

Glossary

Fixed-Pattern Noise

Noise caused by inconsistencies in sensor manufacturing, usually appearing as fixed stripes or spots in the image.

In endoscopic images, fixed-pattern noise affects image clarity and diagnostic accuracy.

Periodic Banding Noise

Noise caused by impedance or phase mismatches in analog signal transmission, usually appearing as periodic stripes in the image.

In endoscopic images, periodic banding noise interferes with image detail and color.

Poisson-Gaussian Noise

Noise caused by random fluctuations in photon arrival and readout noise, usually significant in low-light conditions.

In endoscopic images, Poisson-Gaussian noise affects image contrast and detail.

FPGA

A programmable integrated circuit used for specific computational tasks, offering efficient parallel processing.

In this paper, FPGA is used to achieve real-time image denoising.

PSNR

A metric for measuring image quality, with higher values indicating better quality.

In experiments, PSNR is used to evaluate the effectiveness of the denoising system.

Open Questions Unanswered questions from this research

  • 1 How to further improve system performance in extreme low-light conditions to avoid detail loss?
  • 2 How to enhance system adaptability to different sensor models, reducing calibration steps?

Applications

Immediate Applications

Medical Endoscopes

The system can be directly integrated into existing medical endoscopes, enhancing image quality and diagnostic reliability.

Long-term Vision

Other Medical Imaging Devices

In the future, it can be extended to other types of medical imaging devices, improving overall imaging quality.

Abstract

Endoscopes featuring a miniaturized design have significantly enhanced operational flexibility, portability, and diagnostic capability while substantially reducing the invasiveness of medical procedures. Recently, single-use endoscopes equipped with an ultra-compact analogue image sensor measuring less than 1mm x 1mm bring revolutionary advancements to medical diagnosis. They reduce the structural redundancy and large capital expenditures associated with reusable devices, eliminate the risk of patient infections caused by inadequate disinfection, and alleviate patient suffering. However, the limited photosensitive area results in reduced photon capture per pixel, requiring higher photon sensitivity settings to maintain adequate brightness. In high-contrast medical imaging scenarios, the small-sized sensor exhibits a constrained dynamic range, making it difficult to simultaneously capture details in both highlights and shadows, and additional localized digital gain is required to compensate. Moreover, the simplified circuit design and analog signal transmission introduce additional noise sources. These factors collectively contribute to significant noise issues in processed endoscopic images. In this work, we developed a comprehensive noise model for analog image sensors in medical endoscopes, addressing three primary noise types: fixed-pattern noise, periodic banding noise, and mixed Poisson-Gaussian noise. Building on this analysis, we propose a hybrid denoising system that synergistically combines traditional image processing algorithms with advanced learning-based techniques for captured raw frames from sensors. Experiments demonstrate that our approach effectively reduces image noise without fine detail loss or color distortion, while achieving real-time performance on FPGA platforms and an average PSNR improvement from 21.16 to 33.05 on our test dataset.

eess.IV cs.AI cs.CV